Datex-Ohmeda, Inc. - Avance CS2 and Avance CS2 Pro - Class 1 Recall


Datex-Ohmeda, Inc. - Avance CS2 and Avance CS2 Pro - Class 1 Recall

Potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.GE Healthcare notified consignees on about 07/12/2024 via letter. Consignees were instructed to ensure all potential users complete the Preoperative Checkout procedure in its entirety, as detailed in the URM, make sure that the breathing circuit is correctly connected, only connect an auxiliary manual breathing circuit to the ACGO port, place the included addendum with the URM and inspect all potentially affected devices for the presence of covers applied to the ACGO Port and ACGO Switch. Additionally, customers were requested to complete and return the response form.