If digital angiography system is used with a specific catheterization table and the operator directs simultaneous movement of the table and the C-arm using direct memory, then direct memory button is released, the table may continue to move, which the operator can stop by pressing the stop button.On 9/6/24, recall notices were sent to customers who were asked to do the following: 1) Firmware for the system controller requires updating. The work to correct this issue will be performed by your Authorized Service provider. 2) Schedule time with your Distributer to have your system updated. 3) Complete and return the response form to [email protected] Questions can be directed to the firm's National Technical Support: Toll Free: (844)487-2767 Available 8:00 to 5:00 PM, Monday - Friday